Specifications
Measurement Accuracy Voltage DC Range /Resolution 1 Year 2 Years 100.000 mV 0.02% + 0.005 mV 0.03% + 0.005 mV 3.00000 V 0.02% + 0.00005 V 0.03% + 0.00005 V 30.0000 V 0.02% + 0.0005 V 0.03% + 0.0005 V 300.00 V 0.05% + 0.05 V 0.07% + 0.05 V Voltage AC 3.000 V (40 Hz to 500 Hz) / 0.001 V 0.5% + 0.002 V 1.0% + 0.004 V 30.00 V (40 Hz to 500 Hz) / 0.01 V 0.5% + 0.02 V 1.0% + 0.04 V 300.0 V (40 Hz to 500 Hz) / 0.1 V 0.5% + 0.2 V 1.0% + 0.2 V Current DC 30.000 mA 0.01% + 5 uA 0.015% + 7 uA 110.00 mA 0.01% + 20 uA 0.015% + 30 uA Resistance 10.000 Ω 0.05% + 50 mΩ 0.07% + 70 mΩ 100.00 Ω 0.05% + 50 mΩ 0.07% + 70 mΩ 1.0000 kΩ 0.05% + 500 mΩ 0.07% + 0.5 Ω 10.000 kΩ 0.1% + 10 Ω 0.15% + 15 Ω Frequency 1.00 to 110.00 Hz / 0.01 Hz 0.05 Hz 110.1 to 1100.0 Hz / 0.1 Hz 0.5 Hz 1.101 to 11.000 kHz / 0.001 kHz 0.005 kHz 11.01 to 50.00 kHz / 0.01 kHz 0.05 kHz Source Accuracy 1 Year 2 Years Voltage DC 100.000 mV 0.01% + 0.005 mV 0.015% + 0.005 mV 1.00000 V 0.01% + 0.00005 V 0.015% + 0.0005 V 15.0000 V 0.01% + 0.0005 V 0.015% + 0.0005 V Current DC 22.000 mA (source) 0.01% + 0.003 mA 0.02% + 0.003 mA Current sink (simulate) 0.02% + 0.007 mA 0.04% + 0.007 mA Resistance 10.000 Ω 0.01% + 10 mΩ 0.015% + 15 mΩ 100.00 Ω 0.01% + 20 mΩ 0.015% + 30 mΩ 1.0000 kΩ 0.02% + 0.2 Ω 0.03% + 0.3 Ω 10.000 kΩ 0.02% + 3 Ω 0.03% + 5 Ω Frequency 0.1 to 10.99 Hz 0.01 Hz 0.01 to 10.99 Hz 0.01 Hz 11.00 to 109.99 Hz 0.1 Hz 110.0 to 1099.9 Hz 0.1 Hz 1.100 to 21.999 kHz 0.002 kHz 22.000 to 50.000 kHz 0.005 kHz Technical Data Data log functions Measure functions Voltage, current, resistance, frequency, temperature, pressure Reading rate 1, 2, 5, 10, 20, 30, or 60 readings/minute Maximum record length 8000 readings (7980 for 30 or 60 readings/minute) Ramp functions Source functions Voltage, current, resistance, frequency, temperature Rate 4 steps/second Trip detect Continuity or voltage (continuity detection not available when sourcing current) Loop power function Voltage Selectable, 26 V Accuracy 10%, 18 V minimum at 22 mA Maximum current 25 mA, short circuit protected Maximum input voltage 50 V DC Step functions Source functions Voltage, current, resistance, frequency, temperature Manual step Selectable step, change with arrow buttons Autostep Fully programmable for function, start delay, stepvalue, time per step, repeat Environmental Specifications Operating temperature -10°C to +50°C Storage temperature -20°C to +60°C Dust/water resistance Meets IP52, IEC 529 Operating altitude 3000 m above mean sea level (9842 ft) Safety Specifications Agency approvals CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993 Mechanical and General Specifications Size 136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in) Weight 1.2 kg (2.7 lb) Batteries Internal Battery Pack Li-ion: 7.2 V, 4400 mAh, 30 Wh Battery life > 8 hours typical Battery replacement Via snap-shut door without opening calibrator; no tools required Side port connections Pressure module connector USB Connector to interface to your PC Connection for optional battery charger/eliminator Data storage capacity 1 week of calibration procedures results 90 day specifications The standard specification interval for the 750 Series are 1 and 2 years. Typical 90 day measurement and source accuracy can be estimated by dividing the one year “% of reading” or “% of output” specifications by 2. Floor specifications, expressed as “% of full scale” or “counts” or “ohms” remain constant.