Specifications
Specifications Maximum voltage between any terminal and earth ground 600 V Safety IEC 61010-1: Pollution Degree 2
IEC 61010-2-033: CAT III 600VDisplay Digital: 6,000 counts, updates 4 per second Bar graph 33 segments, updates 32 per second Operating temperature -10°C to + 50°C Storage temperature -40°C to + 60°C Battery type 9 volt Alkaline, IEC 6LR61 Battery life 400 hours typical, without backlight Accuracy specifications DC millivolts Range/resolution 600.0 mV / 0.1 mV Accuracy ± ([% of reading] + [counts]): 0.5% + 2 DC volts Range/resolution 6.000 V / 0.001
V60.00 V / 0.01
V600.00 V / 0.1 VAccuracy ± ([% of reading] + [counts]): 0.5% + 2 Auto volts Range/resolution 600.0 V / 0.1 V Accuracy 2.0% + 3 (DC, 45 Hz to 500 Hz)
4.0% + 3 (500 Hz to 1 kHz)AC millivolts1 True RMS Range/resolution 600.0 mV / 0.1 mV Accuracy 1.0% + 3 (DC, 45 Hz to 500 Hz)
2.0% + 3 (500 Hz to 1 kHz)AC volts1 True RMS Range/resolution 6.000 V / 0.001
V60.00 V / 0.01
V600.0 V / 0.1 VAccuracy 1.0% + 3 (DC, 45 Hz to 500 Hz)
2.0% + 3 (500 Hz to 1 kHz)Continuity Range/resolution 600 Ω / 1 Ω Accuracy Beeper on < 20 Ω, off > 250 Ω detects opens or shorts of
500 μs or longer.Ohms Range/resolution 600.0 Ω / 0.1 Ω
6.000 kΩ / 0.001 kΩ
60.00 kΩ / 0.01 kΩ
600.0 kΩ / 0.1 kΩ
6.000 MΩ / 0.001 MΩAccuracy 0.9% + 1 Range/resolution 40.00 MΩ / 0.01 MΩ Accuracy 5% + 2 1. All ac voltage ranges except Auto-V/LoZ are specified from 1% to 100% of range. Auto-V/LoZ is specified from 0.0 V.
2. Temperature uncertainty (accuracy) does not include the error of the thermocouple probe.
2. Frequency is ac coupled, 5 Hz to 50 kHz for ac voltage. Frequency is dc coupled, 45 Hz to 5 kHz for ac current.
2. Temperature Range/Resolution: -40°F to 752°F / 0.2°F.Mechanical & General Specifications Size 167 x 84 x 46 mm (6.57 x 3.31 x 1.82 in) Weight 550 g