Specifications
General Specifications
Storage temperature
-30°C to 60°C
Operating temperature
-10°C to 55°C
Relative humidity
95% (-10°C to 30°C)
75% (30°C to 40°C)
45% (40°C to 50°C)
35% (50°C to 55°C)
Shock
1 meter drop test
Safety
CSA C22.2 No. 1010.1:1992
EMC
EN50082-1:1992 and EN55022:1994 Class B
Size (H x W x L)
60 x 87 x 210 mm
(2.19 x 3.41 x 8.28 in)
Weight
912 g (2 lb)
Battery
Two 9 V alkaline batteries
Battery life
12 hours with 12 mA into 500 Ω, 200 pump cycles to 100 psi, 500 pump cycles to 30 psi
Warranty
Three-years, one year for pump
Specifications
Fluke-719 30G
Range
-12 to 36 psi, -850 mbar to 2.4 bar
Resolution
0.001 psi, 0.1 mbar
Accuracy
0.025% range for 6 months
0.035% of range for one (1) year
Fluke-719 100G
Range
-12 to 120 psi, -850 mbar to 8 bar
Resolution
0.01 psi, 1 mbar
Accuracy
0.025% range for 6 months
0.035% of range for one (1) year
mA
Range
0 to 24 mA
Resolution
0.001 mA
Accuracy
.015% Rdg + 2 counts
Max load, 1000 Ω, 750 Ω ohm in HART mode
Loop supply
Range
24 V DC
Resolution
N/A
Accuracy
±10%
Pump Specifications
Maximum pressure generation
120 psi (8 BAR) (100 psi model)
Pump to
-30 psi in less than 15 seconds1
-100 psi in less than 45 seconds1
1. With 1 meter of 1/8 inch test hose
Pump cycles between rebuilds
5000
Total pump life
20,000 calibration cycle