Specifications
General Specifications Storage temperature -30°C to 60°C Operating temperature -10°C to 55°C Relative humidity 95% (-10°C to 30°C) 75% (30°C to 40°C) 45% (40°C to 50°C) 35% (50°C to 55°C) Shock 1 meter drop test Safety CSA C22.2 No. 1010.1:1992 EMC EN50082-1:1992 and EN55022:1994 Class B Size (H x W x L) 60 x 87 x 210 mm (2.19 x 3.41 x 8.28 in) Weight 912 g (2 lb) Battery Two 9 V alkaline batteries Battery life 12 hours with 12 mA into 500 Ω, 200 pump cycles to 100 psi, 500 pump cycles to 30 psi Warranty Three-years, one year for pump Specifications Fluke-719 30G Range -12 to 36 psi, -850 mbar to 2.4 bar Resolution 0.001 psi, 0.1 mbar Accuracy 0.025% range for 6 months
0.035% of range for one (1) yearFluke-719 100G Range -12 to 120 psi, -850 mbar to 8 bar Resolution 0.01 psi, 1 mbar Accuracy 0.025% range for 6 months
0.035% of range for one (1) yearmA Range 0 to 24 mA Resolution 0.001 mA Accuracy .015% Rdg + 2 counts Max load, 1000 Ω, 750 Ω ohm in HART mode Loop supply Range 24 V DC Resolution N/A Accuracy ±10% Pump Specifications Maximum pressure generation 120 psi (8 BAR) (100 psi model) Pump to -30 psi in less than 15 seconds1 -100 psi in less than 45 seconds1 1. With 1 meter of 1/8 inch test hose Pump cycles between rebuilds 5000 Total pump life 20,000 calibration cycle