Specifications
Specifications | |||
Measurement Range | 150 mA to 150 A | 100 μΩ to 1 Ω | |
150 mA to 300 A | 10 μΩ to 1 Ω | ||
300 A to 3000 A | 1 μΩ to 1 Ω | ||
Ratio Ranges | 1 to 1,000,000:1 | ||
Linearity | ± 0.01 ppm of full-scale | ||
Warm-Up Time to Full Rated Accuracy | 1 minute | ||
Temperature Coefficient | 0.01 ppm/°C | ||
Communication | IEEE-488 | ||
Test Current Accuracy Note: A 0.1 % current magnitude error is equal to 0.1 ppm in the ratio for supplies that use a reversing switch. | 150 mA to 150 A | ± 0.1 % | |
150 mA to 300 A | ± 0.1 % | ||
300 A to 3000 A | ± 0.1 % of range + 1-bit for 8 hours | ||
Compliance Voltage | 0 to 150 A | 4.2 V | |
150 A to 3000 A | 10 V | ||
Operating Temperature to Full Specification | |||
Maximum Operating Range (30 to 90 % RH) | +10 °C to +40 °C | +50 °F to 104 °F | |
Temperature Storage Range | -20 °C to +60 °C | -4 °F to +140 °F |