Specifications
Mechanical and General Specifications
Size
136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in)
Weight
1.2 kg (2.7 lb)
Batteries
Internal Battery Pack Li-ion: 7.2V,4400mAh, 30 Wh
Battery life
>8 hours typical
Battery replacement
Replace without opening calibrator; no tools required
Side port connections
Pressure module connector
USB Connector to interface to your PC
Digital instrument (HART) connector
Connection for optional battery charger/eliminator
Data storage capacity
1 week of calibration procedures results
90 day specifications
The standard specification interval for the 750 Seriesare 1 and 2 years.
Typical 90 day measurement and source accuracy can be estimated by dividing the one year “% of reading” or “% of output” specifications by 2.
Floor specifications, expressed as “% of full scale” or “counts” or “ohms” remain constant.
Technical Data
Data log functions
Measure functions
Voltage, current, resistance, frequency, temperature, pressure
Reading rate
1, 2, 5, 10, 20, 30, or 60 readings/minute
Maximum record length
8000 readings (7980 for 30 or 60 readings/minute)
Ramp functions
Source functions
Voltage, current, resistance, frequency, temperature
Rate
4 steps/second
Trip detect
Continuity or voltage (continuity detection not available when sourcing current)
Loop power function
Voltage
Selectable, 26 V
Accuracy
10%, 18 V minimum at 22 mA
Maximum current
25 mA, short circuit protected
Maximum input voltage
50 V DC
Step functions
Source functions
Voltage, current, resistance, frequency, temperature
Manual step
Selectable step, change with arrow buttons
Autostep
Fully programmable for function, start delay, stepvalue, time per step, repeat