Specifications
Resistance Measurement 0.1 Ω to 100 kΩ | |||
Range (1:1 Ratio) | Uncertainty (µΩ/Ω) | Range (1:10 Ratio) | Uncertainty (µΩ/Ω) |
0.1 Ω to 0.1 Ω | 0.15 | 0.1 Ω to 1 Ω | 0.1 |
1 Ω to 1 Ω | 0.015 | 1 Ω to 10 Ω | 0.015 |
10 Ω to 10 Ω | 0.015 | 10 Ω to 100 Ω | 0.015 |
100 Ω to 100 Ω | 0.015 | 100 Ω to 1 kΩ | 0.015 |
1 kΩ to 1 kΩ | 0.015 | 1 kΩ to 10 kΩ | 0.015 |
10 kΩ to 10 kΩ | 0.5 | 1 kΩ to 12.9 kΩ | 0.015 |
10 kΩ to 100 kΩ | 0.05 |
- As a ratio device the accuracy specifications can be improved upon based on your standards and environmental conditions.
- Ratio bridge where the ratio accuracies can be verified at anytime using the interchange technique method for 1:1 ratio measurements with the following equation re=(Ra-1/Rb)/2
- Uncertainties follow GUM at 2 sigma level (95%) along with the degrees of freedom
Measurement Mode | 4-wire |
Linearity | < 0.005 x 10-6 of full-scale |
Operating Conditions | 10 °C to 35 °C, 10 % to 90 % RH non-condensing |
Test Current Range | 1 μA to 200 mA |
Test Current Resolution | 18-bit |
Interface | IEEE-488 |
Display | Touch-screen display (no external keyboard), resolution 0.001 x 10-6 |