Specifications
| Model | 7508 | 7511 | 7512 |
| Range | –5 °C to 110 °C | –10 °C to 110 °C | |
| Stability | ±0.0007 °C at 25 °C (water) ±0.001 °C at 25 °C (mineral oil) |
±0.0008 °C at 0 °C (ethanol) ±0.0008 °C at 25 °C (water) ±0.003 °C at 100 °C (oil 5012) |
|
| Uniformity * | ±0.003 °C at 25 °C (water) ±0.004 °C at 25 °C (mineral oil) |
±0.003 °C at 0 °C (ethanol) ±0.003 °C at 25 °C (water) ±0.005 °C at 100 °C (oil 5012) |
|
| Temperature Setting | Digital display with push-button data entry | ||
| Set-Point Resolution | 0.001 °C | ||
| Display Resolution | 0.01 °C or 0.001 °C | ||
| Digital Setting Accuracy | ±1 °C | ||
| Digital Setting Repeatability | ±0.005 °C | ||
| Heaters | 500 and 1000 Watts | ||
| Access Opening (call for customs) |
324 x 184 mm (12.75 x 7.25 in) |
127 x 254 mm (5 x 10 in) |
162 x 292 mm (6.38 x 11.5 in) |
| Depth | 331 mm (13 in) | 305 mm (12 in) | 457 mm (18 in) |
| Wetted Parts | 304 Stainless Steel | ||
| Power | 115 VAC (±10 %), 60 Hz, 12 A (1380 VA); or 230 VAC (±10 %), 50 Hz, 6.39 A (1470 VA); or 208/230 VAC, 60 Hz, 6.39 A (1400 VA) | ||
| Volume | 42 liters (11.2 gal) | 27 liters (7.2 gal) | 42 liters (11.2 gal) |
| Weight | 61 kg (135 lb) | 56.7 kg (125 lb) | 68 kg (150 lb) |
| Size | 610 x 775 x 483 mm (24 x 30.5 x 19 in) |
559 x 686 x 401 mm (22 x 27 x 15.8 in) |
762 x 686 x 401 mm (30 x 27 x 15.8 in) |
| Automation | Use the 1586A Super-DAQ to automate temperature sensor calibration | ||
| Communication | RS-232 + USB A/B | ||
| Display | Color LCD MODULE,TFT,4.3 IN DIAG,480XRGBX272 PIXELS,LED BACKLIGHT | ||
| Refrigerant | R290 Propane (A3 flammable) | ||
| * For 7511 and 7512, working zone definition applies as defined in the user manual | |||