Specifications
Specifications
Current measurement
Range
0 to 24 mA
Resolution
0.001 mA
Accuracy
0.015% Rdg + 2 LSD
Current sourcing
Range
0 to 20 mA or 4 to 20 mA
Accuracy
0.015% Rdg + 2 LSD
Drive capability
1200 Ω 950 Ω at 20 mA in HART mode
Current simulation
Range
0 to 20 mA or 4 to 20 mA
Accuracy
0.015% Rdg + 2 LSD
Ext. loop voltage requirement
12-30 V
Loop power
While measuring mA, 24 V
Voltage measurement
Range
0 to 28 V DC
Resolution
1 mV
Accuracy
0.015% Rdg + 2 LSD
Environmental Specifications
Operating temperature
-10°C to 55°C
Storage temperature
-30°C to 60°C
Humidity (without condensation)
0 to 90% (0°C to 35°C); 0 to 70% (35°C to 55°C)
Temperature coefficient
-10°C to 18°C, 28°C to 55°C, ±.005% of range per °C
Safety Specifications
Safety
CSA C22.2 No. 1010.1:1992
EMC
EN50082-1:1992 and EN55022:1994 Class B
Maximum voltage
30 V
Mechanical and General Specifications
Size
152 x 70 x 41 mm
Weight
224 g (340 g including holster)
Batteries
One 9 V battery Alkaline
Warranty
Three years
Shock and vibration
Per MIL-T-28800 for a Class 2 Instrument
Display current and % of span
Yes
Auto step, auto ramp
Yes
Span check
Yes