Specifications
| General | |
| Power Requirements | 85 to 264 VAC, 50/60 Hz, 25 VA max and 12 VDC @ 9 Ahr |
| Battery/Charger: | 100 to 240 VAC, 50/60 Hz |
| Normal Operating Temperature Range | 15 to 35 °C |
| Vibration | Meets MIL-T-28800D |
| Weight (Typical) | 5 kg (11 lb) |
| Dimensions | 10 cm H x 22.7 cm W x 24 cm D (3.9 in. x 8.9 in. x 9.5 in.) |
| Battery/Charger: | 8 cm H x 22.5 cm W x 20 cm D (3.1 in. x 8.9 in. x 7.9 in.) |
| Communications Ports | RS232 (COM1, COM2), IEEE-488.2 |
| Operating Modes | Absolute, gauge, compound gauge, differential |
| Pressure Ranges | Vacuum to 280 MPa (40,000 psi) |
| Operating Media | Q-RPTs lower than A7M Gas only All others Either gas or oil |
| Calibration | NVLAP accredited |
| Pressure Connections | Up to A70M: 1/8 in. NPT F Above A70M: DH500 (equivalent to AE250C) |
| CE Mark | Available, must be specified |
| Measured pressure (Q-RPT) | |
| Warm Up Time | 30 minute temperature stabilization recommended from cold power up |
| Resolution | To 1 ppm, user adjustable |
| Predicted One Year Stability | ± 0.005 % of reading all ranges and classes |
| Precision | Q-RPTs up to A14M (2,000 psi) |
| Standard Class: | ± 0.008 % of reading, or 0.0024% of Q-RPT span, whichever is greater |
| Premium Class: | ± 0.005 % of reading, 0.0015% of AutoRanged span or 0.0005% of Q-RPT span, whichever is greater |
| Measurement Uncertainty | Q-RPTs up to A12M (2,000 psi) |
| Standard Class: | ± 0.010 % of reading or 0.0030% of Q-RPT span, whichever is greater |
| Premium Class: | ± 0.008 % of reading, 0.0024% of AutoRanged span, or 0.0007% of Q-RPT span, whichever is greater |
| Precision | Q-RPTs A20M TO A140M (3,000 to 20,000 psi) ± 0.012 % of reading or 0.0036% of Q-RPT span, whichever is greater |
| Measurement Uncertainty | Q-RPTs A20M TO A140M (3,000 to 20,000 psi) ± 0.013 % of reading or 0.0039% of Q-RPT span, whichever is greater |
| Precision | QRPT’s A200M to A280M (30,000 to 40,000 psi) ± 0.015 % of reading or 0.0045% of Q-RPT span, whichever is greater |
| Measurement Uncertainty | QRPT’s A200M to A280M (30,000 to 40,000 psi) ± 0.018 % of reading or 0.0054% of Q-RPT span, whichever is greater |